Model Number XOM UNK M-BUR-BLD |
Device Problems
Break (1069); Material Deformation (2976)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/09/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the tool sticks.No patient impact.During analysis of handpiece it was found that the handle (drill) is broken.The base piece of the drill was in the magnum 5 collet.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
H3: analysis confirmed and noted that handle of the drill was broken and stuck inside the m5 handpiece.Visually, there were two broken parts of device returned, the outer and inner hubs.The broken shaft attached to the inner hub measured 0.642 inches from the distal end of the inner hub to the broken point.The distal end of the inner hub was deformed.The outside diameter of the inner hub shall be 0.330 ± 0.002 inches and measured 0.330 inches in undamaged area and up to 0.352 inches in deformed area, which was out of specification.There was no damage noted at the proximal end of the inner hub (seal was intact).The broken part of the outer hub measured 0.632 inches from the proximal end of the outer hub to the broken point.Functional testing could not be performed due to the broken state of the device.In the returned condition, there was an out of specification condition that was related to the complaint (due to physical damage).H6: previous codes fdm b17, fdc c20 and fdr d14 are no longer applicable.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|