Model Number N/A |
Device Problem
Fracture (1260)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/14/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that during the process of disassembling the impactor for cleaning, it was discovered that the impactor tip was broken off inside the handle.There was no patient involvement.
|
|
Event Description
|
No additional information to report.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.H6: proposed component code: mechanical (g04) - impactor the reported event was able to be confirmed via product evaluation.Visual examination of the returned product found it to exhibit signs of repeated use and was nicked, gouged, and worn, and the threaded feature of the component item (# 00133613800) had fractured off in component (# 00133807800).Not all pieces were returned.Fracture surface artifacts of river lines and a bending hinge suggest a likely bending overload fracture.Review of the device history record identified no related deviations or anomalies during manufacturing.The root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|