• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. DRIVER IMPACTOR-FEMORAL; INSTRUMENT, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. DRIVER IMPACTOR-FEMORAL; INSTRUMENT, KNEE Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/14/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that during the process of disassembling the impactor for cleaning, it was discovered that the impactor tip was broken off inside the handle.There was no patient involvement.
 
Event Description
No additional information to report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.H6: proposed component code: mechanical (g04) - impactor the reported event was able to be confirmed via product evaluation.Visual examination of the returned product found it to exhibit signs of repeated use and was nicked, gouged, and worn, and the threaded feature of the component item (# 00133613800) had fractured off in component (# 00133807800).Not all pieces were returned.Fracture surface artifacts of river lines and a bending hinge suggest a likely bending overload fracture.Review of the device history record identified no related deviations or anomalies during manufacturing.The root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRIVER IMPACTOR-FEMORAL
Type of Device
INSTRUMENT, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17480201
MDR Text Key320605661
Report Number0001822565-2023-02096
Device Sequence Number1
Product Code HWR
UDI-Device Identifier00889024186262
UDI-Public(01)00889024186262(10)65477695
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 11/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00133602800
Device Lot Number65477695
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/31/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/17/2023
Initial Date FDA Received08/08/2023
Supplement Dates Manufacturer Received11/06/2023
Supplement Dates FDA Received11/08/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/14/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-