• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD MEDICAL (BD WEST) MEDICAL SURGICAL BD PRE-FILLED SALINE SYRINGE; PISTON SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL (BD WEST) MEDICAL SURGICAL BD PRE-FILLED SALINE SYRINGE; PISTON SYRINGE Back to Search Results
Catalog Number 306594
Device Problem Difficult to Advance (2920)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/25/2023
Event Type  malfunction  
Manufacturer Narrative
H3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that the bd pre-filled saline syringe was damaged and unable to push injection.The following was received by the initial reporter: customers from the box out ready to inject when the front end of the syringe deformation, can not be pushed injection, and this has happened several times in the past, so reported to the hospital spd equipment section,.
 
Manufacturer Narrative
The following fields have been updated due to additional information: d.9.Device available for eval?: yes.D.9.Returned to manufacturer on: 10-aug-2023.H.6.Investigation summary: it was reported the front end of the syringe deformed.To aid in the investigation, one sample with no packaging flow wrap and three photos were provided for evaluation by our quality team.A visual inspection was performed and the syringe barrel luer is bent.The three photos provided show the sample received.No other defects or imperfections were observed.This defect could occur if the unit was not correctly placed in the sterilizing fixture.A device history record review was completed for provided material number 306594, lot 2087541.The review did not reveal any detected quality issues during the production of this lot that could have contributed to the reported defect.There were no related quality notifications.All processes and final inspections complied with specification requirements.The sample will be shown to associates for awareness.Based on the investigation and with the returned sample analysis the symptom reported by the customer is confirmed.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.H3 other text : see h10.
 
Event Description
It was reported that the bd pre-filled saline syringe was damaged and unable to push injection.The following was received by the initial reporter: customers from the box out ready to inject when the front end of the syringe deformation, can not be pushed injection, and this has happened several times in the past, so reported to the hospital spd equipment section.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PRE-FILLED SALINE SYRINGE
Type of Device
PISTON SYRINGE
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer (Section G)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer Contact
jennifer suh
5859 farinon drive
san antonio, TX 78249
8448235433
MDR Report Key17484877
MDR Text Key320921812
Report Number1911916-2023-00566
Device Sequence Number1
Product Code NGT
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number306594
Device Lot Number2087541
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/24/2023
Initial Date FDA Received08/08/2023
Supplement Dates Manufacturer Received09/22/2023
Supplement Dates FDA Received10/03/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/28/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-