During the procedure, prior tachycardia mapping, physician used a (b)(6) needle (medtronic) to perform the transeptal and noted that it was very difficult.The patient became hypotensive and an offset on qrs (segment st) appeared on ecg during the transeptal.The physician decided to wait for the qrs to recover a normal shape before continuing the procedure.After stabilization, mapping was done using the advisor hd grid mapping catheter and ablation was performed with the tacticath ablation catheter.It was clear for the physician (b)(6) devices are not the cause of this issue because he did use an (b)(6) needle for the transeptal.This report reflects information received by fda in the form of a notification per 803.22 (b)(2).
|