• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDIAC SCIENCE CORPORATION POWERHEART G5 KIT, AUTO, US EN/LA ESP, CPRD; AUTOMATED EXTERNAL DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDIAC SCIENCE CORPORATION POWERHEART G5 KIT, AUTO, US EN/LA ESP, CPRD; AUTOMATED EXTERNAL DEFIBRILLATOR Back to Search Results
Model Number G5A-80C
Device Problems High impedance (1291); Unable to Obtain Readings (1516); Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
Complainant alleged that during functional testing, the device failed impedance testing, did not detect electrode pads and the defib output was out of specification.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
 
Manufacturer Narrative
The customer's report was observed during review of the device logs.However, the device was put through extensive testing including daily/weekly/monthly testing and stress testing without duplicating the report.An internal inspection resulted in no findings.The main board was replaced as a precaution.The device was recertified and returned to the customer.The electrode pads used were not returned as part of this investigation.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERHEART G5 KIT, AUTO, US EN/LA ESP, CPRD
Type of Device
AUTOMATED EXTERNAL DEFIBRILLATOR
Manufacturer (Section D)
CARDIAC SCIENCE CORPORATION
500 burdick parkway
deerfield WI 53531 9692
Manufacturer Contact
scott august
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key17491585
MDR Text Key320713182
Report Number2112020-2023-00663
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00812394021222
UDI-Public00812394021222
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P160033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberG5A-80C
Device Catalogue NumberG5A-80C
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/18/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/19/2023
Initial Date FDA Received08/09/2023
Supplement Dates Manufacturer Received07/19/2023
Supplement Dates FDA Received08/29/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/31/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-