Model Number R SERIES
Device Problem Failure to Capture (1081)
Patient Problem Heart Block (4444)
Event Type  malfunction  
Manufacturer Narrative
Zoll medical corporation has received the data file and will be providing a supplemental report when our investigation is completed.
 
Event Description
Complainant alleged that while attempting to pace a patient (age & gender unknown), the device failed to capture the patient's heart rhythm.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
 
Manufacturer Narrative
The device was not returned to zoll medical corporation; instead, the clinical file was received.Review of the clinical file from the reported event concluded no indication of a device malfunction or errors that would indicate the device could not pace.The clinical file review suggests that when the requirements are met, the device was capable of delivering pace therapy.No trend is associated with reports of this type.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R SERIES DEFIBRILLATOR
Type of Device
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
scott august
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key17500068
MDR Text Key320821152
Report Number1220908-2023-03067
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P160022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberR SERIES
Device Catalogue NumberR SERIES
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/21/2023
Initial Date FDA Received08/10/2023
Supplement Dates Manufacturer Received07/21/2023
Supplement Dates FDA Received11/06/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/01/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-