The micor extractor was returned to the manufacturer and evaluated.The device was subjected to visual inspection and functional testing.There was no damage or device malfunction identified and the device met specifications and performed as intended.The published literature was reviewed for comparison of the functional test results for occlusion response; the investigation data shows the device operates within the same expected ranges for similar marketed phacoemulsification devices.The surgical video was provided to the manufacturer for review and the following summary represents the medical review by the company medical affairs director / ophthalmologist.Cataract extraction was completed.As the spatula was removed from the secondary incision aqueous fluid leaked out of the incision and resulted in an unstable anterior chamber.Within 10 seconds, a posterior capsular tear occurred near the 10 - 12 o'clock position.At the time of the tear the micor extractor was positioned near the center cortex and the throttle was still engaged.There was no evidence of anterior chamber collapse and no vitreous fluid loss.The unstable anterior chamber made the capsular bag more susceptible to damage.As with any cataract surgery, caution should be exercised to control aspiration speed when there is evidence of chamber instability.These findings are consistent with the surgeon's assessment that "chamber instability" caused or contributed to the posterior capsule tear.The micor extractor device history records were reviewed for this manufacturing lot and there were no discrepancies or unusual findings that relate to the reported event.Based on the information reviewed, there is no evidence to indicate the presence of a potential quality issue with respect to manufacturing, design, or labeling.The device labeling identifies capsular rupture as an inherent safety risk.Manufacturer's reference #: (b)(4).
|
A patient underwent cataract surgery in the right eye on (b)(6) 2023 where the micor lens fragmentation system (extractor and drive) was used to remove the cataractous lens fragments.Immediately following lens extraction with the micor extractor there was anterior chamber fluctuation (ac shallowing) and the posterior capsule tore.There was no vitreous loss and the planned intraocular lens was implanted.Preoperatively, the patient's best corrected visual acuity (bcva) in the operative eye was 20/50, improving to 20/30 one week postoperatively.At this one-week postoperative visit the patient was reported as stable and improving with 1+ corneal edema that was resolving.
|