ZIMMER BIOMET, INC. G7 VIT E HIGH WALL LNR 36MM D; PROTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Packaging Problem (3007)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/12/2023
Event Type  malfunction  
Event Description
It was reported there was debris in the package.The sterilized box was opened and there was debris that appeared to be a hair in the tray.The product was not used.No known impact or consequence to the patient.
 
Manufacturer Narrative
(b)(4).G2: foreign: japan.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; g3; g6; h1; h2; h3; h6.Visual evaluation of the returned product found a hair-like debris that is not acceptable to specification.The sterile packaging was previously opened.This complaint has been confirmed by evaluation of the returned product.The device history record was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause or condition of the device when it left zimmer biomet cannot be determined, as the product was returned opened.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 VIT E HIGH WALL LNR 36MM D
Type of Device
PROTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17509986
MDR Text Key321242894
Report Number0001822565-2023-02141
Device Sequence Number1
Product Code LPH
UDI-Device Identifier00889024519893
UDI-Public(01)00889024519893(17)280216(10)65833228
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K190660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberN/A
Device Catalogue Number30103604
Device Lot Number65833228
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/02/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/20/2023
Initial Date FDA Received08/11/2023
Supplement Dates Manufacturer Received10/02/2023
Supplement Dates FDA Received10/05/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/17/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-