Model Number N/A |
Device Problem
Packaging Problem (3007)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/12/2023 |
Event Type
malfunction
|
Event Description
|
It was reported there was debris in the package.The sterilized box was opened and there was debris that appeared to be a hair in the tray.The product was not used.No known impact or consequence to the patient.
|
|
Manufacturer Narrative
|
(b)(4).G2: foreign: japan.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; g3; g6; h1; h2; h3; h6.Visual evaluation of the returned product found a hair-like debris that is not acceptable to specification.The sterile packaging was previously opened.This complaint has been confirmed by evaluation of the returned product.The device history record was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause or condition of the device when it left zimmer biomet cannot be determined, as the product was returned opened.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|