The device was returned, and an initial evaluation was conducted by olympus; however, investigation is ongoing.During the initial evaluation, the user¿s report was confirmed.As noted in b5, the angle wire was found broken and compromising the angulation.Additionally, the universal cord was crushed.The video cable protector was cut.The connecting tube was wrinkled and cut, causing a leak.There was also looseness noted at the connection of the bending and insertion sections due to material deformation in the bending tube.If additional information becomes available following the device evaluation, a supplemental report will be filed.
|
The customer reported that the angulation was not adjusting on his olympus cysto-nephro videoscope during procedure preparation.According to the initial reporter, the intended procedure was completed using another device.There was no patient harm reported as a result of this event.During the device evaluation, it was discovered that the angle wire was broken.This report is being submitted to capture the broken angle wire confirmed during the device evaluation.
|