This ra lead was implanted on (b)(6) 2009 and remains implanted at the time.A call was placed to technical services on 10/20/2016 stating that noted atrial intrinsic amplitude measured at 0.2-0.3 mv, impedance stable and put sensitivity down to 0.15 mv, threshold in the a is 5 v at 2 ms, on the egm has biv pacing but expected to see ap if no as is being seen.The physician was dr.(b)(6) at (b)(6) hopital in (b)(6) ca.No other information is availale.This report reflects information received by fda in the form of a notification per 803.22 (b)(2).
|