Joerns healthcare received the included information.The device(s) in question was not manufactured nor imported by joerns healthcare, per section 803.22 (b) (2) we are submitting the following information.Caremed, the manufacturer of the device involved in the incident was notified by joerns on november 29, 2017.The medwatch report (mw5073187) was forwarded to caremed on november 29, 2017.The mattress was found to be at the highest setting in the a.M.Causing the pressure ulcer to get worse.The wound nurse(meg) found the setting in the a.M., tried to change and setting would not change.Technician was called in.Turned the control unit off and on and unit re-set.The pressure ulcer got better after that.The patient is still on the mattress.Injury listed; pressure ulcer got worse.This report reflects information received by fda in the form of a notification per 803.22 (b)(2).
|