MAUDE Adverse Event Report: MEDTRONIC RV LEAD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC RV LEAD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number 6935
Device Problem Protective Measures Problem (3015)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that this right ventricular (rv) lead had a low rv intrinsic amplitude alert.Boston scientific technical services (ts) stated that for that alert any value less than 3.0 mv would trip the alert.There was already a reading of 2.7 mv in the past in the office so suspect the out of range reading was around this value.Lead is operating as programmed.To date, there is no intervention information available and the lead remains in service.No adverse patient effects were reported.This report reflects information received by fda in the form of a notification per 803.22 (b)(2).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RV LEAD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
MEDTRONIC
MDR Report Key17534539
MDR Text Key321274596
Report NumberMW5141441
Device Sequence Number1
Product Code LWS
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Other
Type of Report Initial
Report Date 12/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Model Number6935
Initial Date Manufacturer Received Not provided
Initial Date FDA Received08/12/2023
Patient Sequence Number1
-
-