• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. CYSTO-NEPHRO VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. CYSTO-NEPHRO VIDEOSCOPE Back to Search Results
Model Number CYF-VHA
Device Problem Deformation Due to Compressive Stress (2889)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The customer reported to olympus that the cysto-nephro videoscope had an air leak.Upon inspection and evaluation, scraping of the forceps stopper cap.There was no report of patient harm or user injury associated with this event.This medical device report (mdr) is being submitted to capture the reportable malfunction found during incoming inspection and evaluation.
 
Manufacturer Narrative
The device was returned to olympus for inspection, and the customer's issue of ¿air leak" was confirmed.The following findings were noted during device evaluation: insufficient bending angle, resin float, insertion tube collapse, scratches and wrinkles, universal cord crushed and scratched, video cable crushed and scratched, rattling at the connection between the curved tube and the tip, scratches on curved rubber joints, video connector scratches, light guide connector scratches, video connector case scratches, operation part scratch, switch box scratch, operation part cover scratch, grip scratch, up/down scratch, up/down plate scratch, up/down angle fixing lever scratch, and vent metal misalignment.The investigation is ongoing, and follow-up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the instrument channel port was confirmed to be scraped.However, the definitive root cause of the scraped instrument channel port could not be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CYSTO-NEPHRO VIDEOSCOPE
Type of Device
CYSTO-NEPHRO VIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17538096
MDR Text Key321628877
Report Number3002808148-2023-08316
Device Sequence Number1
Product Code FAJ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K062049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 08/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCYF-VHA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/18/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/18/2023
Initial Date FDA Received08/13/2023
Supplement Dates Manufacturer Received08/21/2023
Supplement Dates FDA Received08/25/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/27/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-