• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND IN.PACT ADMIRAL PACLITAXEL-ELUTING PTA BALLOON CATHETER; DRUG-ELUTING PERIPHERAL TRANSLUMINAL ANGIOPLAS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND IN.PACT ADMIRAL PACLITAXEL-ELUTING PTA BALLOON CATHETER; DRUG-ELUTING PERIPHERAL TRANSLUMINAL ANGIOPLAS Back to Search Results
Catalog Number ADM05008013P
Device Problem Burst Container or Vessel (1074)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/01/2023
Event Type  malfunction  
Event Description
Physician was attempting to use inpact admiral paclitaxel coated balloon catheter with a 65cm 6fr non-medtronic sheath and a 0.035 non-medtronic guidewire during procedure to treat a calcified lesion in the patients left proximal popliteal artery.Little vessel tortuosity and moderate vessel calcification are reported.Embolic protection was not used.A non-medtronic inflation device was used for balloon inflation.Contrast and saline were used for balloon inflation.There was no damage noted to packaging, i.E.Shelf carton, hoop/tray.No issues were noted when removing the device from the hoop/tray.Ifu was followed and the device was prepped without issue.The device did not pass through a previously deployed stent.No resistance was noted during advancement of the device and no excessive force was used.It was reported that during balloon inflation, a balloon burst occurred at 12 atm.There were no prior in flations applied to balloon prior to issue.Balloon did not fragment, no intervention required to remove.A replacement inpact admiral dcb was used to complete procedure.No vessel damage noted.No patient injury reported.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IN.PACT ADMIRAL PACLITAXEL-ELUTING PTA BALLOON CATHETER
Type of Device
DRUG-ELUTING PERIPHERAL TRANSLUMINAL ANGIOPLAS
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key17547704
MDR Text Key321144476
Report Number9612164-2023-03685
Device Sequence Number1
Product Code ONU
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140010
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberADM05008013P
Device Lot Number0011372394
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/01/2023
Initial Date FDA Received08/15/2023
Supplement Dates Manufacturer Received08/30/2023
Supplement Dates FDA Received09/29/2023
Date Device Manufactured08/25/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexMale
-
-