Investigation was completed.The instrument log files of the reported system and two comparative systems were reviewed and analyzed.The summary log determines that one na_cal threshold dead error was detected by system about 5 hours before first reported sample measurement followed by 8 sodium sensor interferent detected events, 14 d2 excessive drifts, and 3 d3 slope errors for the na+ sensor.Consequently, the na+ sensor was disabled several times by the system to maintain normal functioning.The errors indicate the sensor was exposed to sodium sensor interferent (unknown source) several times which, as stated in the system manual, makes the sodium sensor unstable during the time when reported samples were measured.According to the information provided, the rp500e instrument is operating normally at the customer site.
|