It was reported that the balloon would not deflate, causing vessel trauma upon removal.This 7.0 mm x 40 mm, 135 cm ranger balloon was selected for use in a percutaneous transluminal angioplasty procedure of an arm dialysis fistula.During the procedure, the balloon was inflated for three minutes in the lesion.Balloon deflation was attempted several times without success.Therefore, the balloon was removed while inflated with the guidewire and sheath, causing vessel trauma.It was further noted that this created a rather large arteriotomy at the access site.Manual pressure was applied to the site for a prolonged amount of time.The patient fully recovered.
|