Catalog Number CLXUSA
Device Problem Coagulation in Device or Device Ingredient (1096)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/20/2023
Event Type  malfunction  
Event Description
The customer contacted mallinckrodt to report blood clotting in the return line with their cellex photopheresis kit ("kit") during an extracorporeal photopheresis (ecp) treatment.The customer reported they received an alarm #53: return line air detected alarm during the procedure when approximately 1493 ml of whole blood had been processed.The customer stated they observed platelet aggregates in the blood filter.The customer inspected the kit and noted there were blood clots in the return bag and return line.The ecp treatment was aborted and residual blood within the kit was not returned to the patient.The customer reported the patient was in stable condition.The customer did not return product for evaluation.
 
Manufacturer Narrative
The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction clot observed in the return line.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.A batch record review for kit lot m303 was conducted.There were no non-conformances associated with this lot.This lot met all release requirements.A review of kit lot m303 shows no trends.Trends were reviewed for complaint categories, alarm #53: return line air detected and clot observed.No trends were detected for these complaint categories.No product was returned for evaluation.Section 2-9 of the cellex operators manual (1470493 rev 6) for use with software 5.4 on anticoagulant states "caution: individual patients may require a heparin dosage that varies from the recommended dose to prevent post-treatment bleeding or clotting during a treatment.The physician should review the patient's medical condition, medications and platelet count at the time of treatment and use clinical judgement to establish the optimal heparin dosage for each patient." the root cause for the clotting observed could not be determined based on the available information.No further action is required at this time.This investigation is now complete.(b)(4).H.M.(b)(6) 2023.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELLEX PHOTOPHERESIS SYSTEM
Type of Device
CELLEX PHOTOPHERESIS SYSTEM
Manufacturer (Section D)
MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED
blanchardstown, dublin, D15 T X2V
EI  D15 TX2V
Manufacturer (Section G)
MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED
college business & tech park
cruiserath road
blanchardstown, dublin, D15 T X2V
EI   D15 TX2V
Manufacturer Contact
mark wendelken
440 us rt 22 east
suite 302
bridgewater, NJ 08807
MDR Report Key17552303
MDR Text Key321726271
Report Number3013428851-2023-00045
Device Sequence Number1
Product Code LNR
UDI-Device Identifier20705030200003
UDI-Public(01)20705030200003(10)M303(17)250101
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P860003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 08/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCLXUSA
Device Lot NumberM303
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/20/2023
Initial Date FDA Received08/15/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/17/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age42 YR
Patient SexMale
-
-