• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ISOFLEX OPTIM LEAD; PERMANENT DEFIBRILLATOR ELECTRODES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ISOFLEX OPTIM LEAD; PERMANENT DEFIBRILLATOR ELECTRODES Back to Search Results
Model Number 1944/52
Device Problems Failure to Capture (1081); Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/10/2023
Event Type  malfunction  
Event Description
It was reported that the patient presented for an implant procedure.During the procedure, the atrial lead exhibited failure to capture and failure to sense.The atrial lead was not used and replaced.The patient was in stable condition.
 
Manufacturer Narrative
The reported events of failure to capture and failure to sense were not confirmed.A complete lead was returned in one piece.Electrical testing, visual inspection and x-ray examinations of the lead were normal with no anomalies found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ISOFLEX OPTIM LEAD
Type of Device
PERMANENT DEFIBRILLATOR ELECTRODES
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17554214
MDR Text Key321242893
Report Number2017865-2023-38449
Device Sequence Number1
Product Code NVY
UDI-Device Identifier05414734502320
UDI-Public05414734502320
Combination Product (y/n)Y
Reporter Country CodeIN
PMA/PMN Number
P960030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1944/52
Device Lot NumberP000155948
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/12/2023
Initial Date FDA Received08/16/2023
Supplement Dates Manufacturer Received11/15/2023
Supplement Dates FDA Received11/22/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/17/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-