Model Number LTF-S190-5
Device Problem Image Display Error/Artifact (1304)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/27/2023
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation and the customer's allegation was confirmed.The device evaluation found that there was no image, only black/grey lines and the video connector needed troubleshooting to recheck the image.The scope passed a leakage test and all switches were noted to be working.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
The customer reported the image of the endoeye flex deflectable videoscope would pop up and then abruptly cut out.The event occurred during the preparation for use.There were no reports of patient or user harm associated with this event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root case of the observed no image issue could not be determined, however, it is likely that the issue was the result of a faulty printed circuit board and or video connector.The event can be detected by following the instructions for use section below: -inspection of the endoscopic image.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE
Type of Device
FLEX DEFLECTABLE VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17555821
MDR Text Key321771616
Report Number9610595-2023-11736
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier04953170310355
UDI-Public04953170310355
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201832
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 09/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLTF-S190-5
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/31/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/31/2023
Initial Date FDA Received08/16/2023
Supplement Dates Manufacturer Received09/01/2023
Supplement Dates FDA Received09/12/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/27/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-