• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL ADVISOR¿ HD GRID MAPPING CATHETER, SENSOR ENABLED¿; ELECTRODE RECORDING CATHETER OR ELECTRODE RECORDING PROBE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL ADVISOR¿ HD GRID MAPPING CATHETER, SENSOR ENABLED¿; ELECTRODE RECORDING CATHETER OR ELECTRODE RECORDING PROBE Back to Search Results
Model Number D-AVHD-DF16
Device Problem Contamination /Decontamination Problem (2895)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/24/2023
Event Type  malfunction  
Event Description
During an atrial flutter procedure, a white clear thread like material was wrapped around the paddle at the kinked spline when the mapping was completed, and the catheter was removed from the patient.It was noted that one of the catheter splines was kinked.The catheter was not reintroduced into the patient.The procedure was completed and there were no adverse consequences to the patient.
 
Manufacturer Narrative
Additional information revealed the initial mdr for this event was reported under the incorrect mfr report # and manufacturing site.Corrected manufacturing site information has been provided in d3.The correct mfr number is 3005334138.
 
Manufacturer Narrative
One bi-directional, curve d-f, sensor enabled, advisor hd grid mapping catheter was received for evaluation.The catheter had a foreign material wrapped around the distal coupler cap.The s&t report identified the foreign material as ptfe (poly tetraflouroethylene).The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the foreign material remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVISOR¿ HD GRID MAPPING CATHETER, SENSOR ENABLED¿
Type of Device
ELECTRODE RECORDING CATHETER OR ELECTRODE RECORDING PROBE
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
edificio #44b, calle 0, avenida 2, coyol
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL
5050 nathan lane north
edificio #44b, calle 0, avenida 2, coyol
plymouth MN 55442
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key17557012
MDR Text Key321350915
Report Number3008452825-2023-00353
Device Sequence Number1
Product Code DRF
UDI-Device Identifier05415067028198
UDI-Public05415067028198
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K172393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 11/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberD-AVHD-DF16
Device Lot Number9041540
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/08/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/25/2023
Initial Date FDA Received08/16/2023
Supplement Dates Manufacturer Received10/25/2023
11/02/2023
Supplement Dates FDA Received10/25/2023
11/03/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/11/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-