• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. DS2ADV AUTO CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. DS2ADV AUTO CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number DSX520T11C
Device Problem Gas/Air Leak (2946)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/29/2021
Event Type  malfunction  
Event Description
The manufacturer received information alleging the dreamstation 2 advanced auto cpap device air is leaking.There was no report of patient harm or injury.The device was returned to the manufacturer's product investigation laboratory for evaluation.In evaluation it is find that the device is having burn marks on the heater plate.Ran device for 15 minutes with 15mm heated tube and there was no air leak detected.Device pressure running around 12 cmh2o.Streamweaver data shows a consistent pressure.E200 error present in error log (looks to be from connecting usb module).After taking the device apart, the iso port was loose and the screw was able to be tightened.Assuming assembly error, along with the patients positioning of the device, caused the tubing to pull on the iso port allowing air to leak the device forwarded to philips product investigations lab, failure mode confirmed and unit subsequently.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DS2ADV AUTO CPAP
Type of Device
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key17557369
MDR Text Key321291211
Report Number2518422-2023-19893
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K200480
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberDSX520T11C
Device Catalogue NumberDSX520T11C
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/14/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/29/2021
Initial Date FDA Received08/16/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/22/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-