Model Number LTF-S190-5 |
Device Problem
Image Display Error/Artifact (1304)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device was returned as part of the asset return process and in addition to the reported in b5, the device evaluation found that the bending section cover rubber had a scratch, the adhesive on the bending section cover rubber had a chip, and the venting connector of the light guide connector was loose.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided.
|
|
Event Description
|
The endoeye flex deflectable videoscope was returned as part of the asset return process.During routine inspection of the device by olympus, it was observed that due to damage on the charged coupled device unit the image noise occurred, and the image cannot be seen temporarily.There was no procedural or patient involvement associated with this event.This mdr is being submitted to capture the reportable malfunction found during the device evaluation.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the event occurred due to breakage or disconnection of the image sensor unit due to stress of repeated use, external factors or handling, or failure of the parts mounted on the electrical circuit board such as integrated circuit chips and capacitors.The event can be prevented by following the instructions for use (ifu) which state: "chapter 3 preparation and inspection, section 3.8 inspection of the endoscopic system.[inspection of the endoscopic image] confirm that the endoscopic image is displayed normally in wli (white light imaging) and nbi (narrow band imaging) observation.1 before inspection, wipe the objective lens using clean lint-free cloths moistened with saline solution or sterilized water.2 observe the palm of your hand in the wli and nbi endoscopic images.3 confirm that light is output from the endoscope¿s distal end.4 adjust the brightness level as appropriate.5 confirm that the endoscopic image is free from noise, blur, fog, or other irregularities during wli and nbi observation.6 turn the angulation control levers slowly in each direction until it stops.7 confirm that the wli and nbi endoscopic images do not momentarily disappear or display any other irregularities." olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|
|
|