A male patient underwent an aquablation procedure for symptomatic benign prostatic hyperplasia (bph).Procept biorobotics corporation became aware that during the aquablation procedure, the aquabeam handpiece high-velocity waterjet was not pumping water at the planned treatment power, although the aquabeam handpiece had been successfully primed during setup.A second handpiece was used and the aquablation procedure was successfully completed.However, it was reported that the treating surgeon had to resect more tissue than initially planned due to this issue.The reported event caused a surgical delay of over 20 minutes.There were no adverse health consequences to the patient due to this event.
|
H.3 device evaluation by manufacturer: the aquabeam handpiece was returned for investigation.The reported event could not be confirmed as three docking cycles and priming at 50%/100% pump power were performed successfully.No issues were observed before, during, or after the treatment phase, and a simulated aquablation session was performed successfully.The handpiece flow rates were measured at 25%, 50%, 75%, and 100% pump power, and the values were within the required range.Thus, the handpiece functioned as designed.Additionally, the handpiece probe tip was observed under magnification.No physical damage, obstruction, or blockage could be seen that would affect jetting.However, the root cause remains undeterminable.A review of the device history record (dhr) ab2000-b/ serial number (b)(6) and the aquabeam handpiece/lot number 23c00903 was conducted, which confirmed that there were no non-conformances, failures, discrepancies, or missed steps during the manufacturing process that could be related to the reported event.The review indicated that the system met all design and manufacturing specifications when released for distribution.The current instruction for use ifu0101-00 rev.E, aquabeam robotic system ifu, us, english was reviewed.8.30 sterile: treatment a.To begin the aquablation treatment, step on the foot pedal and gently support the beige braided tubing extending from the back of the aquabeam handpiece.Caution (for system models other than ab2000c): failure to support the beige braided tubing may result in a system fault or insufficient cutting efficacy.A.During the aquablation treatment, use [-] indicator on the motorpack to decrease power/resection depth (if needed) and use [+] to increase back to planned power submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
|