• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PROCEPT BIOROBOTICS CORPORATION AQUABEAM ROBOTIC SYSTEM; FLUID JET REMOVAL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PROCEPT BIOROBOTICS CORPORATION AQUABEAM ROBOTIC SYSTEM; FLUID JET REMOVAL SYSTEM Back to Search Results
Catalog Number AB2000
Device Problem Infusion or Flow Problem (2964)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/24/2023
Event Type  malfunction  
Manufacturer Narrative
Root cause of reported event has not yet been established.Investigation by manufacturer is currently in-process.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Event Description
A male patient underwent an aquablation procedure for symptomatic benign prostatic hyperplasia (bph).Procept biorobotics corporation became aware that during the aquablation procedure, the aquabeam handpiece high-velocity waterjet was not pumping water at the planned treatment power, although the aquabeam handpiece had been successfully primed during setup.A second handpiece was used and the aquablation procedure was successfully completed.However, it was reported that the treating surgeon had to resect more tissue than initially planned due to this issue.The reported event caused a surgical delay of over 20 minutes.There were no adverse health consequences to the patient due to this event.
 
Manufacturer Narrative
H.3 device evaluation by manufacturer: the aquabeam handpiece was returned for investigation.The reported event could not be confirmed as three docking cycles and priming at 50%/100% pump power were performed successfully.No issues were observed before, during, or after the treatment phase, and a simulated aquablation session was performed successfully.The handpiece flow rates were measured at 25%, 50%, 75%, and 100% pump power, and the values were within the required range.Thus, the handpiece functioned as designed.Additionally, the handpiece probe tip was observed under magnification.No physical damage, obstruction, or blockage could be seen that would affect jetting.However, the root cause remains undeterminable.A review of the device history record (dhr) ab2000-b/ serial number (b)(6) and the aquabeam handpiece/lot number 23c00903 was conducted, which confirmed that there were no non-conformances, failures, discrepancies, or missed steps during the manufacturing process that could be related to the reported event.The review indicated that the system met all design and manufacturing specifications when released for distribution.The current instruction for use ifu0101-00 rev.E, aquabeam robotic system ifu, us, english was reviewed.8.30 sterile: treatment a.To begin the aquablation treatment, step on the foot pedal and gently support the beige braided tubing extending from the back of the aquabeam handpiece.Caution (for system models other than ab2000c): failure to support the beige braided tubing may result in a system fault or insufficient cutting efficacy.A.During the aquablation treatment, use [-] indicator on the motorpack to decrease power/resection depth (if needed) and use [+] to increase back to planned power submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AQUABEAM ROBOTIC SYSTEM
Type of Device
FLUID JET REMOVAL SYSTEM
Manufacturer (Section D)
PROCEPT BIOROBOTICS CORPORATION
150 baytech dr.
suite 101
san jose CA 95134
Manufacturer (Section G)
PROCEPT BIOROBOTICS CORPORATION
150 baytech dr.
suite 101
san jose CA 95134
Manufacturer Contact
doria esquivel
150 baytech dr.
suite 101
san jose, CA 95134
6502327291
MDR Report Key17560513
MDR Text Key321811592
Report Number3012977056-2023-00139
Device Sequence Number1
Product Code PZP
UDI-Device IdentifierB614AB20001
UDI-Public+B614AB20001/16D202101084
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
DEN170024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberAB2000
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/25/2023
Initial Date FDA Received08/16/2023
Supplement Dates Manufacturer Received10/22/2023
Supplement Dates FDA Received10/30/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/08/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexMale
-
-