Model Number LTF-S190-10
Device Problem Image Display Error/Artifact (1304)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/19/2023
Event Type  malfunction  
Event Description
The customer reported to olympus, the endoeye flex deflectable videoscope had an image abnormality during angulation.The issue was found during preparation for an unspecified therapeutic procedure.The intended procedure was completed using a similar device.There was no report of patient or user harm associated with the event.During testing and inspection of the returned device, it was discovered that the charge coupled device (ccd) unit in the endoscope connector, the image tone was abnormal (magenta).This report is being submitted to capture this reportable malfunction which was identified during the evaluation.
 
Manufacturer Narrative
During the evaluation, the specified resolving power was not obtained, error code e311 was displayed, and there was image noise, due to the damaged ccd unit.The distal end cover had a scratch, switch buttons 1 and 3, and the video cable cover, had scratches.The video connector had a crack.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the reportable events occurred since the image sensor cable was kinked within the bending section of the insertion section.It is likely that the cause of the kink of the image sensor cable was excessive twisting, bending, or other stress that exceeded expectations.The event can be prevented by following the instructions for use (ifu) which state: "chapter 3 preparation and inspection 3.8 inspection of the endoscopic system in the operation manual.[inspection of the endoscope] confirm that the wli (white light imaging) and nbi (narrow band imaging) endoscopic images are normal.1 before inspection, wipe the objective lens using clean lint-free cloths moistened with saline solution or sterilized water.2 observe the palm of your hand in the wli and nbi endoscopic images.3 confirm that light is output from the endoscope¿s distal end.4 adjust the brightness level as appropriate.5 confirm that the wli and nbi endoscopic images are free from noise, blur, fog, or other irregularities.6 turn the angulation control levers slowly in each direction until it stops.7 confirm that the wli and nbi endoscopic images do not momentarily disappear or display any other irregularities." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE
Type of Device
FLEX DEFLECTABLE VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17560979
MDR Text Key321348958
Report Number9610595-2023-11786
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier04953170310386
UDI-Public04953170310386
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K111425
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 10/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLTF-S190-10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/25/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/28/2023
Initial Date FDA Received08/16/2023
Supplement Dates Manufacturer Received10/06/2023
Supplement Dates FDA Received10/12/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/21/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse