• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. UNIVERSAL INSERTER/EXTRACTOR; INSTRUMENT, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. UNIVERSAL INSERTER/EXTRACTOR; INSTRUMENT, HIP Back to Search Results
Model Number N/A
Device Problems Fracture (1260); Difficult to Remove (1528)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/26/2023
Event Type  malfunction  
Event Description
It was reported that the stem inserter was cracked and the black plastic is spinning making it difficult to remove from the stem.The handle was eventually removed by grabbing the lower part.Nothing was noticed falling off the handle.The surgery was completed without complication.There is no additional information available at the time of this report.
 
Manufacturer Narrative
(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.H3 other text : device in process to return.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Corrected: d4: lot number.
 
Event Description
There is no update to the prior event description provided.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual examination of the returned product identified the black grip of the handle to be cracked.While holding the shaft and handle in the lab, the technician was unable to recreate the spinning motion of the handle alleged in the complaint description.The shaft is heavily scuffed and dinged.The thread flutes have been rounded off.The strike plate also exhibits scratching and dings.Review of the device history records identified no deviations or anomalies during manufacturing.Medical records were not provided.A definitive root cause cannot be determined if any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
There is no update to the prior event description provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSAL INSERTER/EXTRACTOR
Type of Device
INSTRUMENT, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17568187
MDR Text Key321451678
Report Number0001825034-2023-01945
Device Sequence Number1
Product Code HWB
UDI-Device Identifier05019279457403
UDI-Public(01)05019279457403(11)180222(10)ZB180101
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 11/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number31-473601
Device Lot NumberZB180801
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/26/2023
Initial Date FDA Received08/17/2023
Supplement Dates Manufacturer Received08/18/2023
11/20/2023
Supplement Dates FDA Received08/24/2023
11/29/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/22/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
UNKNOWN STEM
Patient SexPrefer Not To Disclose
-
-