• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAYLIS MEDICAL COMPANY INC. VERSACROSS CONNECT LAAC ACCESS SOLUTION; DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAYLIS MEDICAL COMPANY INC. VERSACROSS CONNECT LAAC ACCESS SOLUTION; DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION Back to Search Results
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Low Blood Pressure/ Hypotension (1914); Cardiac Tamponade (2226); Diaphoresis (2452); Pericardial Effusion (3271)
Event Date 07/26/2023
Event Type  Injury  
Event Description
It was reported that during a watchman left atrial appendage closure (laac) procedure, a pericardial effusion and cardiac tamponade occurred post procedure.After obtaining right femoral vein access, an nrg rf transseptal kit (nrg needle and troflex sheath) was inserted.There were multiple attempts of dropping onto the fossa and transseptal puncture was performed however while crossing with stiff wire, the right superior pulmonary vein (rspv) access was lost.The physician then performed another transseptal puncture using versacross connect kit (pigtail rf wire and single curve sheath/dilator) with no complications.The sheath was advanced into the left atrial appendage (laa) and the dilator was removed.Act result was 230 and an additional 2000 units of heparin given.Watchman device was inserted with one recapture performed, met the pass criteria and was released.All devices were removed from the body.There was no effusion noted at this point.After one hour while the patient being in recovery, patient became hypotensive and diaphoretic.It was discovered that the patient had a pericardial effusion with cardiac tamponade.The patient was brought back to the lab where the physician performed a pericardiocentesis.675ml of blood was removed from the patient before the patient became stable.The patient was admitted to hospital beyond standard of care and expected to be fully recovered.The device is not expected to be returned for analysis.
 
Manufacturer Narrative
It was indicated that the device will not be returned for evaluation.If there is any further relevant information obtained, a supplemental medwatch will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSACROSS CONNECT LAAC ACCESS SOLUTION
Type of Device
DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION
Manufacturer (Section D)
BAYLIS MEDICAL COMPANY INC.
5959 trans-canada highway
montreal H4T 1 A1
CA  H4T 1A1
Manufacturer (Section G)
BAYLIS MEDICAL COMPANY INC.
5825 explorer drive
mississauga, on L4W 5 P6
CA   L4W 5P6
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330
saint paul, MN 55112
6515826168
MDR Report Key17568575
MDR Text Key321402520
Report Number2124215-2023-43958
Device Sequence Number1
Product Code DRE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K220414
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 08/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/26/2023
Initial Date FDA Received08/17/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Required Intervention; Hospitalization;
Patient Age79 YR
Patient RaceWhite
-
-