Model Number OER-4
Device Problem Inaccurate Flow Rate (1249)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/26/2023
Event Type  malfunction  
Event Description
The field service engineer reported to olympus, the water flow in the endoscope reprocessor was weak.The issue occurred during reprocessing.There were no reports of patient harm.
 
Manufacturer Narrative
The device was not returned to olympus for evaluation as the field service engineer was on site.The device evaluation found that there were two reprocessors with the same issue and found that the filter in both reprocessors were not changed since august 7, 2019.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the definitive root cause of the water filter issue could not be determined.It is possible that the user did not thoroughly read the instruction manual and mismanaged the water filter's replacement.The instruction manual identifies the following verbiage, which may have prevented the phenomenon: ¿replace the water filter periodically, at least once a month to prevent contamination of the rinse water.The water filter should also be replaced whenever the error code [e01] indicating water supply insufficiency is displayed.Replace the water filter according to the procedure described below.¿ olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPE REPROCESSOR
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17568937
MDR Text Key321805935
Report Number9610595-2023-11862
Device Sequence Number1
Product Code FEB
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 09/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOER-4
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/26/2023
Initial Date FDA Received08/17/2023
Supplement Dates Manufacturer Received08/29/2023
Supplement Dates FDA Received09/05/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/01/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-