The national support specialist (nss) visited the customers facility to investigate the issue.The nss performed an inspection of the medical devices at the facility.Visual inspection found that multiple devices had broken battery retention tabs, corrosion on lead sets, and improper cleaning damage.The following functional tests were performed: the nss captured logs from a time the issue occurred.The nss also states that he was able to recreate the issue.The nss confirms that the when the device is connected to the network, the appropriate alarms are sent and acknowledged.The nss also confirms that the customer was notified via alarm when the devices lost connectivity.The nss confirms the device damage as the cause of the issue.Good faith effort was performed to obtain additional device serial and part numbers however, this information was not made available.Based on the information available and the testing conducted, the cause of the reported problem was damage to the mx40 device resulting in intermittent ecg waveforms when alarms occurred.The reported problem was confirmed.He customer is working with philips sales team to make a bulk order replacement device purchase.It has been concluded that no further action is required at this time.If additional information is received the complaint file will be reopened.
|