• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON AND CO. BD POSIFLUSH¿ NORMAL SALINE SYRINGE; SALINE, VASCULAR ACCESS FLUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON AND CO. BD POSIFLUSH¿ NORMAL SALINE SYRINGE; SALINE, VASCULAR ACCESS FLUSH Back to Search Results
Catalog Number 306572
Device Problems Contamination (1120); Contamination /Decontamination Problem (2895)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/03/2023
Event Type  malfunction  
Event Description
It was reported that the bd posiflush¿ normal saline syringe experienced foreign matter, contaminated.The following information was provided by the initial reporter, translated from german to english: there are black particles on the syringe plunger.Foreign bodies or impurities can enter the bloodstream!.
 
Manufacturer Narrative
E.1 initial reporter phone #: (b)(6).H3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Manufacturer Narrative
To aid in the investigation of this issue, one (1) syringe sample was returned for evaluation by our quality team.Through examination of the sample, the syringe/plunger was found to have embedded foreign matter.This type of defect most likely occurred during the molding process.A device history record review was completed for provided material number 306572 and lot number 2208047.There was one possible non-conformance identified which could have contributed to this reported incident.The non-conformance was related to threads on the plunger rod, which was discovered during the startup checks.The manufacturing technician was notified, and the affected product and manufacturing components were purged accordingly.Based on the investigation conclusions, it is likely that the molding machine was not purged for long enough, resulting in contaminated material.At this time, further action has not been determined necessary, as this is the first report received for this type of defect for this lot.Our quality team will continue to closely monitor the manufacturing process for signs of this potential defect and any emerging trends.
 
Event Description
No additional information.There are black particles on the syringe plunger.Foreign bodies or impurities can enter the bloodstream.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD POSIFLUSH¿ NORMAL SALINE SYRINGE
Type of Device
SALINE, VASCULAR ACCESS FLUSH
Manufacturer (Section D)
BECTON, DICKINSON AND CO.
donore road
drogheda
Manufacturer (Section G)
BECTON, DICKINSON AND CO.
donore road
drogheda
Manufacturer Contact
jennifer suh
5859 farinon drive
san antonio, TX 78249
8448235433
MDR Report Key17577826
MDR Text Key321632755
Report Number9616657-2023-00034
Device Sequence Number1
Product Code NGT
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number306572
Device Lot Number2208047
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/04/2023
Initial Date FDA Received08/18/2023
Supplement Dates Manufacturer Received11/06/2023
Supplement Dates FDA Received11/13/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/27/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-