The subject device was returned to olympus for evaluation.During inspection and testing, the allegation was confirmed.It was not possible to inject liquid when the slider was pushed.However, it was possible to inject liquid when the slider was pulled.The needle could extend from the outer tube.The outer tube was not buckled.The needle tube was presented with compressive buckling.A review of the device history record (dhr) found no deviations that could have caused or contributed to the reported malfunction.The dhr confirmed that the subject device was shipped in accordance with the specifications.It is determined that liquid could not be injected due to compressive buckling of the needle tube.Since friction resistance between the outer tube and needle tube had increased, compressive buckling might have occurred when the needle was extended.It is presumed the friction resistance between the outer tube and needle tube occurred due to the following mechanisms: - the tube was coiled during the inspection for operation.- the slider was pushed abruptly.The instruction manual provides the following: ¿·straighten out the instrument before inspecting it.The instrument can be damaged if it is coiled while the handle is operated.·operate the slider slowly, otherwise the tube could buckle.·do not coil the insertion portion with a diameter of less than 15 cm.This could damage the insertion portion.·when inserting the instrument into the endoscope, retract the needle into the sheath, hold the instrument close to the biopsy valve, and keep it as straight as possible relative to the biopsy valve.Otherwise, the instrument could be damaged.·stop using the instrument if the insertion portion bends excessively during use.This could result in malfunction, such as failing to extend the needle or inject a fluid.¿ olympus will continue to monitor field performance for this device.
|
An olympus employee reported on behalf of a customer, the single use injector needle was pulled out, the syringe could not be pushed, and no water could out.The event occurred during a therapeutic endoscopic submucosal dissection (esd).The device was completed using a replacement device.There was no report of patient harm associated with this event.
|