• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIO-DETEK INCORPORATED PRO-PADZ RADIOLUCENT SOLID GEL ELECTRODE, SINGLE; DEFIB/PACING ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIO-DETEK INCORPORATED PRO-PADZ RADIOLUCENT SOLID GEL ELECTRODE, SINGLE; DEFIB/PACING ELECTRODE Back to Search Results
Model Number 8900-4006
Device Problem Arcing of Electrodes (2289)
Patient Problem Atrial Fibrillation (1729)
Event Date 07/27/2023
Event Type  malfunction  
Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
 
Event Description
Complainant alleged that while attempting to cardiovert an 84-year-old female patient, an arc (loud pop) was heard from the electrode pad.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
 
Manufacturer Narrative
Zoll medical corporation evaluated the electrode pads and the customer's report was not replicated or confirmed.Visual inspection of the electrode pads observed some slight discoloration around part of the outer edges of the tin of the front pad.There were no signs of any burns or sparking.The pads passed continuity testing.The pads were scrapped.No clinical data was provided for review.During defibrillation therapy, sparks could occur due to a high measured patient impedance.This could be due to many reasons including, but not limited to, how the skin is prepared, how the electrodes are applied, or poor coupling.Electrode labeling provides instructions for proper electrode application technique.Poor adherence and /or air under the electrodes can lead to the possibility of arcing.It is likely that there is poor coupling between the pads and the patient, but without the device log or additional information regarding patient prep or pads application, root cause cannot be determined.No trend is associated with reports of this type.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRO-PADZ RADIOLUCENT SOLID GEL ELECTRODE, SINGLE
Type of Device
DEFIB/PACING ELECTRODE
Manufacturer (Section D)
BIO-DETEK INCORPORATED
525 narragansett park drive
pawtucket RI 02861
Manufacturer Contact
tina lombari
525 narragansett park drive
pawtucket, RI 02861
4017291400
MDR Report Key17584591
MDR Text Key321541042
Report Number1218058-2023-00132
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00847946016661
UDI-Public00847946016661
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date04/30/2024
Device Model Number8900-4006
Device Catalogue Number8900-4006
Device Lot Number1823
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/10/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/31/2023
Initial Date FDA Received08/21/2023
Supplement Dates Manufacturer Received07/31/2023
Supplement Dates FDA Received09/19/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/30/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age84 YR
Patient SexFemale
-
-