• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ISOFLEX OPTIM LEAD; PERMANENT DEFIBRILLATOR ELECTRODES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ISOFLEX OPTIM LEAD; PERMANENT DEFIBRILLATOR ELECTRODES Back to Search Results
Model Number 1948/58
Device Problems Failure to Capture (1081); Failure to Sense (1559); Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/31/2023
Event Type  Injury  
Event Description
It was reported that the patient presented one day post implant with loss of capture and failure to sense exhibited by the right ventricular lead.It was confirmed that the lead had dislodged.The lead was explanted and replaced.The patient was stable.
 
Manufacturer Narrative
The reported events were lead dislodgement, loss of capture and failure to sense.As received, a complete lead was returned in one piece with all four tines were found to be intact and undamaged.The reported events of loss of capture and sensing were not confirmed.Electrical tests did not find any indication of conductor fractures or internal shorts.Visual and x-ray inspections of the lead did not find any anomalies.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ISOFLEX OPTIM LEAD
Type of Device
PERMANENT DEFIBRILLATOR ELECTRODES
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17587050
MDR Text Key321569924
Report Number2017865-2023-38899
Device Sequence Number1
Product Code NVY
UDI-Device Identifier05414734502351
UDI-Public05414734502351
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P960030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1948/58
Device Lot NumberA000141348
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/07/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/31/2023
Initial Date FDA Received08/21/2023
Supplement Dates Manufacturer Received08/28/2023
Supplement Dates FDA Received09/18/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/24/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ASSURITY; TENDRIL STS
Patient Outcome(s) Required Intervention;
Patient SexMale
-
-