Catalog Number 07212771190 |
Device Problems
High Test Results (2457); Low Test Results (2458); Non Reproducible Results (4029)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/12/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The field service engineer (fse) performed routine maintenance and service.The fse noted that the liquid flow cleaning (lfc) system was working fine.He performed mechanical and precision checks with successful results.The investigation excluded a general reagent problem.Based on the information provided, the investigation did not identify a product problem.The cause of the event could not be determined.
|
|
Event Description
|
The initial reporter received questionable elecsys vitamin b12 immunoassay results from two patient samples tested on the cobas e411 disk.The reporter was able to provide one example of a questionable result.The initial result was greater than 2000 pg/ml.The repeat result was 225.2 pg/ml.The repeat result was cross-checked on the laboratory's other e 411 and no significant variation was noted.
|
|