Catalog Number 07212771190
Device Problems High Test Results (2457); Low Test Results (2458); Non Reproducible Results (4029)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/12/2023
Event Type  malfunction  
Manufacturer Narrative
The field service engineer (fse) performed routine maintenance and service.The fse noted that the liquid flow cleaning (lfc) system was working fine.He performed mechanical and precision checks with successful results.The investigation excluded a general reagent problem.Based on the information provided, the investigation did not identify a product problem.The cause of the event could not be determined.
 
Event Description
The initial reporter received questionable elecsys vitamin b12 immunoassay results from two patient samples tested on the cobas e411 disk.The reporter was able to provide one example of a questionable result.The initial result was greater than 2000 pg/ml.The repeat result was 225.2 pg/ml.The repeat result was cross-checked on the laboratory's other e 411 and no significant variation was noted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS VITAMIN B12 II
Type of Device
RADIOASSAY, VITAMIN B12
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key17591119
MDR Text Key321838805
Report Number1823260-2023-02717
Device Sequence Number1
Product Code CDD
Combination Product (y/n)Y
Reporter Country CodeIN
PMA/PMN Number
K060755
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number07212771190
Device Lot Number707331
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/26/2023
Initial Date FDA Received08/21/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-