Model Number 866389 |
Device Problem
Alarm Not Visible (1022)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/25/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
Philips is in process of obtaining additional information.A final report will be submitted upon completion of the investigation.
|
|
Event Description
|
The customer reported that unspecified patient harm occurred related to alarm type not showing at the central.
|
|
Manufacturer Narrative
|
Visual inspection found this was a onetime incident per the field service engineer.The customer has not reported any further incidents.Based on the information available and the testing conducted we were unable to replicate the reported problem.The reported problem was not confirmed.The device was confirmed to be operating per specifications and no failure was identified.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
|
|