RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60
Device Problem No Audible Alarm (1019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/15/2023
Event Type  malfunction  
Event Description
Philips received a complaint on the v60 ventilator indicating that there was a master alarm failure.The issue was identified outside of clinical use and no reports of user harm or injury.The customer informed the authorized service provider (asp) that they observed the message display on the monitor.The asp recommended to the customer to replace the speaker assembly.This investigation is ongoing.
 
Manufacturer Narrative
Additional information received where it was reported that parts were ordered, and the device was repaired.It confirmed that the replacement parts resolved the reported issue.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Manufacturer Narrative
H10: additional information received from the authorized service provider (asp) that clarified that it was actually the central processing unit (cpu) printed circuit board assembly (pcba) which was replaced, not the speaker assembly.The asp also stated that the device issue occurred during testing of the device.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key17591806
MDR Text Key321652084
Report Number2518422-2023-20301
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838025776
UDI-Public00884838025776
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberV60
Device Catalogue Number1076716
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/15/2023
Initial Date FDA Received08/21/2023
Supplement Dates Manufacturer Received08/22/2023
10/18/2023
Supplement Dates FDA Received08/25/2023
10/25/2023
Date Device Manufactured03/16/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-