A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the event occurred due to the following (user handling): uhi-4 was installed lower than the patient.Uhi-4 was connected to an insufflation tube, and the tube was installed higher than the co2 feed port while blood or body fluids were inside the tube.The relief mode was turned on without using a filter.The event can be prevented by following the instructions for use (ifu) which state: "[chapter 5 operation] install this instrument at a higher altitude than the patient.Otherwise, the blood and body fluid of the patient may contaminate the high-flow insufflation unit through the insufflation tube.[5.12 steps after use] if blood and body fluid penetrate inside the insufflation tube that is connected to the high-flow insufflation unit, be careful not to bring the insufflation tube at a higher level than the co2 feed port of the insufflation unit when disconnecting the tube.Otherwise, the blood and body fluid may enter the insufflation unit and contaminate it.[5.13 relief mode] when relief mode is set to on, the intra-cavity gas will flow backward into the equipment for relief through the built-in valve.To prevent the internal surfaces of the equipment from being contaminated, be sure to use a disposable filter.When no filter is used, make sure relief mode is set to off." olympus will continue to monitor field performance for this device.
|