• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERPORT SLIM IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, KIT, 6F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERPORT SLIM IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, KIT, 6F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Catalog Number 1716070J
Device Problems Fracture (1260); Deformation Due to Compressive Stress (2889); Material Deformation (2976); Material Integrity Problem (2978)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 07/26/2023
Event Type  Injury  
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us, but is similar to the powerport slim implantable port, chronoflex single-lumen, kit, 6f products that are cleared in the us.The pro code and 510k number for the powerport slim implantable port, chronoflex single-lumen, kit, 6f products is identified in d2 and g4.H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.H10: d4 (expiration date: 09/2024).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that during a port placement procedure, the tip of the dilator was allegedly damaged.Reportedly, patient experienced tissue damaged.The current status of the patient is reported to be stable.
 
Event Description
It was reported that during a port placement procedure, the tip of the dilator was allegedly damaged.Reportedly, patient experienced tissue damaged.The current status of the patient is reported to be stable.
 
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us, but is similar to the powerport slim implantable port, chronoflex single-lumen, kit, 6f products that are cleared in the us.The pro code and 510k number for the powerport slim implantable port, chronoflex single-lumen, kit, 6f products is identified in d2 and g4.H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one vessel dilator was returned for evaluation.Gross visual and microscopic evaluation were performed.The vessel dilator was noted to be bent and the distal tip of the vessel dilator was noted to be jagged and fractured.Therefore, the investigation is confirmed for the reported deformation due to compressive stress and identified fracture issue.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiration date: 09/2024), g3.H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us, but is similar to the powerport slim implantable port, chronoflex single-lumen, kit, 6f products that are cleared in the us.The pro code and 510k number for the powerport slim implantable port, chronoflex single-lumen, kit, 6f products is identified in d2 and g4.H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one vessel dilator was returned for evaluation.Gross visual and microscopic evaluation were performed.The distal tip of the vessel dilator was noted to be jagged and fractured.Therefore, the investigation is confirmed for the reported dilator damage and fracture issue was identified upon evaluation.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiration date: 09/2024).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that during a port placement procedure, the tip of the dilator was allegedly damaged.Reportedly, patient experienced tissue damaged.The current status of the patient is reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPORT SLIM IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, KIT, 6F
Type of Device
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key17594693
MDR Text Key321625239
Report Number3006260740-2023-03604
Device Sequence Number1
Product Code LJT
UDI-Device Identifier00801741163739
UDI-Public(01)00801741163739
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K072549
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 04/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number1716070J
Device Lot NumberREHN2159
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/26/2023
Initial Date FDA Received08/22/2023
Supplement Dates Manufacturer Received10/26/2023
05/17/2024
Supplement Dates FDA Received11/06/2023
05/27/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/24/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-