Catalog Number 221261 |
Device Problem
Device Contamination with Chemical or Other Material (2944)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/13/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
E.6 initial reporter e-mail: (b)(6).H.3.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
|
Event Description
|
It was reported that while using the bd bbl¿ trypticase soy agar with 5% sheep blood (tsa ii) that there was biological contamination.The following information was provided by the initial reporter: customer is reporting contaminated media for product 221261 lot no.3122794.
|
|
Manufacturer Narrative
|
H.6.Investigation summary: during manufacturing of material 221261, media is formulated and sent through a high temperature short time sterilizer to remove bioburden.The petri dishes are subjected to uv radiation to decrease bioburden.The petri dishes are filled in a positive pressure hepa filtered environment.The filled plates are cooled and immediately wrapped into sleeves to decrease the introduction of microbes.Sleeves are then packaged into cartons and then transferred to a refrigerated truck (2 to 8 degrees c) for shipment to the distributor.Bd distributors are provided with the storage guidelines for the shipping and handling of bd media of 2 to 8 degrees c in a dark place.The batch history record for batch 3122794 was reviewed and no quality notifications were generated during manufacturing and inspection.The release testing that is performed on this product does include bioburden testing.A sample of plates are incubated at 25 degrees c and at 35 degrees c for approximately 72 hours.All bioburden testing performed on this batch was satisfactory per bd internal procedures.Affected product does not have any sterility claims; the product is tested for bioburden prior to release to ensure that it conforms to product specifications.However, this does not ensure that the end-user will not receive a contaminated plate.The complaint history for batch 3122794 was reviewed and no other complaints have been taken.No retention samples for batch 3122794 were available for inspection.Four photos were received for investigation of this complaint.The first photo shows a plate (batch number and time stamp visible) with contamination on the agar surface.The second photo shows a plate from batch 3122794 (time stamp 0400) with contamination.The third photo shows a plate from batch 3122794 (time stamp 0400) with contamination.The fourth photo shows a plate (batch number and time stamp not visible) with contamination on the agar surface.No return samples were received for investigation.This complaint can be confirmed.H3 other text : see h.10.
|
|
Event Description
|
It was reported that while using the bd bbl¿ trypticase soy agar with 5% sheep blood (tsa ii) that there was biological contamination.The following information was provided by the initial reporter: customer is reporting contaminated media for product 221261 lot no.3122794.
|
|
Search Alerts/Recalls
|