BECTON, DICKINSON & CO. (SPARKS) BD BBL¿ TRYPTICASE SOY AGAR WITH 5% SHEEP BLOOD (TSA II); CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL Back to Search Results
Catalog Number 221261
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/13/2023
Event Type  malfunction  
Manufacturer Narrative
E.6 initial reporter e-mail: (b)(6).H.3.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that while using the bd bbl¿ trypticase soy agar with 5% sheep blood (tsa ii) that there was biological contamination.The following information was provided by the initial reporter: customer is reporting contaminated media for product 221261 lot no.3122794.
 
Manufacturer Narrative
H.6.Investigation summary: during manufacturing of material 221261, media is formulated and sent through a high temperature short time sterilizer to remove bioburden.The petri dishes are subjected to uv radiation to decrease bioburden.The petri dishes are filled in a positive pressure hepa filtered environment.The filled plates are cooled and immediately wrapped into sleeves to decrease the introduction of microbes.Sleeves are then packaged into cartons and then transferred to a refrigerated truck (2 to 8 degrees c) for shipment to the distributor.Bd distributors are provided with the storage guidelines for the shipping and handling of bd media of 2 to 8 degrees c in a dark place.The batch history record for batch 3122794 was reviewed and no quality notifications were generated during manufacturing and inspection.The release testing that is performed on this product does include bioburden testing.A sample of plates are incubated at 25 degrees c and at 35 degrees c for approximately 72 hours.All bioburden testing performed on this batch was satisfactory per bd internal procedures.Affected product does not have any sterility claims; the product is tested for bioburden prior to release to ensure that it conforms to product specifications.However, this does not ensure that the end-user will not receive a contaminated plate.The complaint history for batch 3122794 was reviewed and no other complaints have been taken.No retention samples for batch 3122794 were available for inspection.Four photos were received for investigation of this complaint.The first photo shows a plate (batch number and time stamp visible) with contamination on the agar surface.The second photo shows a plate from batch 3122794 (time stamp 0400) with contamination.The third photo shows a plate from batch 3122794 (time stamp 0400) with contamination.The fourth photo shows a plate (batch number and time stamp not visible) with contamination on the agar surface.No return samples were received for investigation.This complaint can be confirmed.H3 other text : see h.10.
 
Event Description
It was reported that while using the bd bbl¿ trypticase soy agar with 5% sheep blood (tsa ii) that there was biological contamination.The following information was provided by the initial reporter: customer is reporting contaminated media for product 221261 lot no.3122794.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD BBL¿ TRYPTICASE SOY AGAR WITH 5% SHEEP BLOOD (TSA II)
Type of Device
CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL
Manufacturer (Section D)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer (Section G)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer Contact
jennifer suh
5859 farinon drive
san antonio, TX 78249
8448235433
MDR Report Key17605418
MDR Text Key321765550
Report Number1119779-2023-00922
Device Sequence Number1
Product Code JSG
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
PREAMENDMENT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date08/29/2023
Device Catalogue Number221261
Device Lot Number3122794
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/13/2023
Initial Date FDA Received08/23/2023
Supplement Dates Manufacturer Received10/20/2023
Supplement Dates FDA Received10/24/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/02/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-