MEDTRONIC PUERTO RICO OPERATIONS CO. SURESCAN; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR Back to Search Results
Model Number 97702
Device Problems Break (1069); Difficult to Remove (1528); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Device Embedded In Tissue or Plaque (3165); Insufficient Information (4580)
Event Date 07/31/2023
Event Type  Injury  
Manufacturer Narrative
Concomitant medical products: product id 3093-28; lot# v243906; implanted: (b)(6) 2009; explanted: (b)(6) 2009.Product type lead product id 3093-28; lot# v243906; implanted: (b)(6) 2009; explanted: (b)(6) 2009.Product type lead section d information references the main component of the system.Other relevant device(s) are: product id: 3093-28, serial/lot #: (b)(6), ubd: 21-apr-2013, udi#: (b)(4); product id: 3093-28, serial/lot #: (b)(6), ubd: 21-apr-2013, udi#: (b)(4).D2/g4: please note that this device was used in an off-label manner, as it was implanted for urinary dysfunction/sacral nerve stim, overactive bladder, and urgency incontinence medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from an healthcare provider regarding a patient who was implanted with an implantable neurostimulator (ins) for urinary dysfunction/sacral nerve stim, overactive bladder, and urgency incontinence.It was reported that the patient was getting less stimulation and was requiring high amounts of stimulation form the device.During the removal procedure, one of the leads was frayed and it was attempted to remove the leads, but there was absolutely no movement.Given the caudal placement, it would have been difficult or dangerous to do anything more aggressive to remove the leads, so they were cut at the sacrum and left intact along the nerve.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESCAN
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key17624442
MDR Text Key321962793
Report Number3004209178-2023-14914
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00643169974883
UDI-Public00643169974883
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 08/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/28/2020
Device Model Number97702
Device Catalogue Number97702
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/31/2023
Initial Date FDA Received08/25/2023
Date Device Manufactured06/13/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
"SEE H10...."
Patient Outcome(s) Required Intervention;
Patient Age75 YR
Patient SexFemale
-
-