Concomitant medical products: product id 3093-28; lot# v243906; implanted: (b)(6) 2009; explanted: (b)(6) 2009.Product type lead product id 3093-28; lot# v243906; implanted: (b)(6) 2009; explanted: (b)(6) 2009.Product type lead section d information references the main component of the system.Other relevant device(s) are: product id: 3093-28, serial/lot #: (b)(6), ubd: 21-apr-2013, udi#: (b)(4); product id: 3093-28, serial/lot #: (b)(6), ubd: 21-apr-2013, udi#: (b)(4).D2/g4: please note that this device was used in an off-label manner, as it was implanted for urinary dysfunction/sacral nerve stim, overactive bladder, and urgency incontinence medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
Information was received from an healthcare provider regarding a patient who was implanted with an implantable neurostimulator (ins) for urinary dysfunction/sacral nerve stim, overactive bladder, and urgency incontinence.It was reported that the patient was getting less stimulation and was requiring high amounts of stimulation form the device.During the removal procedure, one of the leads was frayed and it was attempted to remove the leads, but there was absolutely no movement.Given the caudal placement, it would have been difficult or dangerous to do anything more aggressive to remove the leads, so they were cut at the sacrum and left intact along the nerve.
|