There was an allegation of questionable gen.2 ise indirect for cl results for 3 patient samples on a cobas integra 400 plus.The customer was prompted to repeat the samples as their afternoon qc was out of range, the results did not meet their clinical pictures, and they questioned the high results.On (b)(6) 2023, sample id (b)(6) had an initial cl result of 146 mmol/l.The repeat result was 104 mmol/l.On (b)(6) 2023, sample id (b)(6) had an initial cl result of 185 mmol/l.The repeat result was 107 mmol/l.On (b)(6), sample id (b)(6) had an initial cl result of 139 mmol/l.The repeat result was 104 mmol/l.No questionable results were reported outside of the laboratory.The repeat results were deemed correct.
|