The device was returned to olympus for a device evaluation and the customer¿s allegation was not confirmed.In addition to the reported in b5, the device evaluation found the following: the distal end had a dent, due to damage on the ccd unit the image had white dots, the control unit had a scratch, the grip had a scratch, the up/down plate had a scratch, the right/left plate had a scratch, the light guide connector had a scratch, the light guide cover glass had a scratch, switch 1 had a scratch, the right/left lever had a scratch, and the video connector had a scratch.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided.
|
The customer reported to olympus that the endoeye flex deflectable videoscope had a noisy image.The issue was observed during preparation for use on a therapeutic procedure.The procedure was completed with a similar device and there were no reports of patient or user harm associated with this event.The device was returned to olympus for a device evaluation and found due to damage on the ccd (charged coupled device) unit, image noise occurs and the image could not be seen and due to damage on the ccd unit, a b31 (scope communication error) occurred.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
|