• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE Back to Search Results
Model Number LTF-S190-10
Device Problems Image Display Error/Artifact (1304); Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for a device evaluation and the customer¿s allegation was not confirmed.In addition to the reported in b5, the device evaluation found the following: the distal end had a dent, due to damage on the ccd unit the image had white dots, the control unit had a scratch, the grip had a scratch, the up/down plate had a scratch, the right/left plate had a scratch, the light guide connector had a scratch, the light guide cover glass had a scratch, switch 1 had a scratch, the right/left lever had a scratch, and the video connector had a scratch.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided.
 
Event Description
The customer reported to olympus that the endoeye flex deflectable videoscope had a noisy image.The issue was observed during preparation for use on a therapeutic procedure.The procedure was completed with a similar device and there were no reports of patient or user harm associated with this event.The device was returned to olympus for a device evaluation and found due to damage on the ccd (charged coupled device) unit, image noise occurs and the image could not be seen and due to damage on the ccd unit, a b31 (scope communication error) occurred.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE
Type of Device
FLEX DEFLECTABLE VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17628715
MDR Text Key322031525
Report Number9610595-2023-12258
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier04953170310386
UDI-Public04953170310386
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K111425
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 08/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLTF-S190-10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/07/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/07/2023
Initial Date FDA Received08/25/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/25/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-