Model Number PM2272
Device Problem Inappropriate or Unexpected Reset (2959)
Patient Problem Syncope/Fainting (4411)
Event Date 08/07/2023
Event Type  Injury  
Event Description
It was reported, that patient presented at the emergency room, symptomatic of syncope.Initial transmission of the device noted, that the pacemaker was in backup mode, due to hardware reset.The subsequent transmission revealed, that the device was functioning normally.Interrogation in clinic noted, normal device function.No interventions were performed.The patient was in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSURITY MRI
Type of Device
Implantable pulse generator, pacemaker (non-CRT)
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17628990
MDR Text Key322019861
Report Number2017865-2023-39381
Device Sequence Number1
Product Code LWP
UDI-Device Identifier05414734509589
UDI-Public05414734509589
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 08/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2022
Device Model NumberPM2272
Device Lot NumberA000100670
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/07/2023
Initial Date FDA Received08/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/03/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MRI LEAD; TENDRIL LEAD
Patient Outcome(s) Other;
Patient Age84 YR
Patient SexFemale
Patient Weight80 KG
-
-