• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS GOLDWAY (SHENZHEN) INDUSTRIAL INC. EFFICIA DFM100; DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS GOLDWAY (SHENZHEN) INDUSTRIAL INC. EFFICIA DFM100; DEFIBRILLATOR Back to Search Results
Model Number 866199
Device Problems Material Integrity Problem (2978); Device Fell (4014)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/03/2023
Event Type  malfunction  
Event Description
It was reported to philips that the efficia dfm100 suffered a fall.There was no reported patient involvement.The efficia dfm100 defibrillator, model# 866199, is substantially similar to the heartstart xl+ defibrillator (model # 861290) and will be reported in the united states under device model # 861290.
 
Manufacturer Narrative
Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.
 
Manufacturer Narrative
This report is based on information provided by philips field service engineer (fse) personnel and has been investigated by the philips complaint handling team regarding a complaint received about the efficia dfm100, which indicated that the equipment suffered a fall.There was reportedly no patient involvement.Visual inspection and functional testing were conducted by the field service engineer (fse) to assess the equipment's condition after a fall, determining that only the blade connector fastener was broken and recommending its replacement, with a quote to be sent for the necessary service repair.Based on the information available and the testing conducted, the root cause of the reported incident, following numerous tests, has been identified as the sole failure component within the equipment subsequent to its fall: the blade connector fastener.As a result, to ensure the equipment's optimal functionality, it will be imperative to replace this specific part.However, the factors leading to the equipment's fall remain unknown at this time.Based on the information available and results of additional analysis, no further action is necessary at this time.An analysis was performed by a vigilance reporting specialist (vrs).The vrs determined that while no harm was alleged, recurrence of this failure mode during clinical use could cause or contribute to death or serious injury.Therefore, this failure mode meets the definition of a reportable malfunction.Appropriate regulatory reporting actions have been taken.The data entered in this complaint record will be utilized for product quality and safety improvements per the post market surveillance and risk management processes.The resolution was to send a quote for replacing the damaged part, specifically the blade connector fastener, to ensure the equipment's perfect operation.The investigation concludes with no further action required at this time, but the complaint file will be reopened if additional information is received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EFFICIA DFM100
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS GOLDWAY (SHENZHEN) INDUSTRIAL INC.
no.2 keji north 3rd road
nanshan district
shenzhen WA
CH 
Manufacturer (Section G)
PHILIPS GOLDWAY (SHENZHEN) INDUSTRIAL INC.
no.2 keji north 3rd road
nanshan district
shenzhen WA
CH  
Manufacturer Contact
feng she
no.2 keji north 3rd road
nanshan district
shenzhen 
CH  
7552698099
MDR Report Key17632732
MDR Text Key322072053
Report Number3030677-2023-03384
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
K110825
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number866199
Device Catalogue Number866199
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/03/2023
Initial Date FDA Received08/28/2023
Supplement Dates Manufacturer Received08/03/2023
Supplement Dates FDA Received10/17/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/30/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-