• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EPIC VASCULAR; STENT, ILIAC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EPIC VASCULAR; STENT, ILIAC Back to Search Results
Model Number 39200
Device Problems Difficult to Remove (1528); Positioning Problem (3009)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/08/2023
Event Type  Injury  
Event Description
It was reported that the stent was difficult to position.The target lesion was located in iliac artery.A 10x40x75 epic stent was advanced for treatment.However, upon initial deployment of the stent, the stent jumped forward and deeper into the aorta than intended.The physician then tried to pull back the stent but with more than half of the stent being deployed it could not be pulled back or repositioned.Hence, the physician decided to place another stent in ostium iliac lesion to help preserve the iliac bifurcation.Moreover, the stent was fully opposed to the vessel wall post-procedure.No patient complications were reported, and the patient fully recovered.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIC VASCULAR
Type of Device
STENT, ILIAC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key17637296
MDR Text Key322111637
Report Number2124215-2023-44728
Device Sequence Number1
Product Code NIO
UDI-Device Identifier08714729805014
UDI-Public08714729805014
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P110035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 08/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number39200
Device Catalogue Number39200
Device Lot Number0030933654
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/08/2023
Initial Date FDA Received08/28/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/27/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-