BAXTER HEALTHCARE CORPORATION PRISMAX MACHINES; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 955626
Device Problems Protective Measures Problem (3015); Unexpected Shutdown (4019)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 08/05/2023
Event Type  Injury  
Event Description
It was reported that, during continuous renal replacement therapy with a prismax machine, the machine "ended treatment with call service powering off suddenly without restarting itself".The bedside nurse reported treatment was ended without the extracorporeal blood being returned to the patient.It was further reported the patient received a blood transfusion.No additional information is available.
 
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H10: a service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.The device was not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
Additional information added to h6 and h10.H10: the event history log was further reviewed.During review the log showed the system halted operation on (b)(6) 2023 at 21:33 due to a b1897 ¿ replacement scale load cell mismatch.The replacement bag was just changed.The reported condition was not verified; however, the most probable cause was due to the act of closing the scale.This could not be verified without a device analysis.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAX MACHINES
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - BROOKLYN PARK
7601 northland drive
n suite 170
minneapolis MN 55428
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key17641844
MDR Text Key322168970
Report Number3003504604-2023-00022
Device Sequence Number1
Product Code KDI
UDI-Device Identifier00085412639499
UDI-Public(01)00085412639499
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K190910
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
Report Date 12/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number955626
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/05/2023
Initial Date FDA Received08/29/2023
Supplement Dates Manufacturer Received10/09/2023
11/08/2023
Supplement Dates FDA Received10/31/2023
12/06/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-