Catalog Number 955626 |
Device Problems
Protective Measures Problem (3015); Unexpected Shutdown (4019)
|
Patient Problem
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 08/05/2023 |
Event Type
Injury
|
Event Description
|
It was reported that, during continuous renal replacement therapy with a prismax machine, the machine "ended treatment with call service powering off suddenly without restarting itself".The bedside nurse reported treatment was ended without the extracorporeal blood being returned to the patient.It was further reported the patient received a blood transfusion.No additional information is available.
|
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
H10: a service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.The device was not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
Additional information added to h6 and h10.H10: the event history log was further reviewed.During review the log showed the system halted operation on (b)(6) 2023 at 21:33 due to a b1897 ¿ replacement scale load cell mismatch.The replacement bag was just changed.The reported condition was not verified; however, the most probable cause was due to the act of closing the scale.This could not be verified without a device analysis.Should additional relevant information become available, a supplemental report will be submitted.
|
|