Catalog Number 105200-000030 |
Device Problem
Material Separation (1562)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/15/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Qn#(b)(4).Other remarks: n/a.Corrected data: n/a.
|
|
Event Description
|
It is reported that "as the anesthesiologist was removing the lma at the end of the procedure, the ventilation tube detached from the cuff leaving the cuff in place.The anesthesiologist was able to extract the cuff by the inflation port tubing.Please note that the patient was not resisting or biting on the lma, so this seemed to be a random event".No patient harm or injury.The patient status is reported as "fine".
|
|
Manufacturer Narrative
|
Qn# (b)(4).The reported complaint of "detached - cuff" was confirmed based upon the sample received.The customer returned one unit of 105200-000030 lma unique silicone cuff pilot size 3.Visual examination of the returned sample revealed that glue was observed at the joint surface and on the airway tube surface joint area.The returned sample appeared used as there was biological material present on the device.A device history record review was performed, and no relevant findings were identified.The root cause of the complaint is considered as a "manufacturing related" issue as per the returned sample.Therefore, this complaint was concluded as confirmed.An investigation was initiated to further investigate the issue.The investigation is still on-going.Teleflex will continue to monitor and trend on complaints of this nature.
|
|
Event Description
|
It is reported that "as the anesthesiologist was removing the lma at the end of the procedure, the ventilation tube detached from the cuff leaving the cuff in place.The anesthesiologist was able to extract the cuff by the inflation port tubing.Please note that the patient was not resisting or biting on the lma, so this seemed to be a random event".No patient harm or injury.The patient status is reported as "fine".
|
|
Search Alerts/Recalls
|