• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. COMP AUG MINI BSPLT W TPR SM; EXTREMITIES IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. COMP AUG MINI BSPLT W TPR SM; EXTREMITIES IMPLANT Back to Search Results
Model Number N/A
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Hypersensitivity/Allergic reaction (1907); Pain (1994); Metal Related Pathology (4530)
Event Date 07/31/2023
Event Type  Injury  
Manufacturer Narrative
(b)(4).D10/d11: concomitant medical products - medical product: catalog #: 115310, comp rvrs shldr glnsp std 36mm, lot # 263010.Catalog #: 115370, comp rvs tray co 44mm, lot # 403720.Catalog #: 180553, comp lk scr 3.5hex 4.75x30 st, lot # 606680.Catalog #: 180552, comp lk scr 3.5hex 4.75x25 st, lot # 504390.Catalog #: 180552, comp lk scr 3.5hex 4.75x25 st, lot # 427570.Catalog #: 113631, comp primary stem 11mm mini, lot # 733660.Catalog #: 115396, comp rvs cntrl 6.5x30mm st/rst, lot # 919910.Catalog #: ep-115393, e1 44-36 std hmrl brng, lot # 961790.H3: customer has indicated that the product will not be returned to zimmer biomet for investigation, as the has been requested but not returned by hospital.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2023-02027, 0001825034-2023-02029, 0001825034-2023-02030, 0001825034-2023-02031, 0001825034-2023-02032, 0001825034-2023-02033, 0001825034-2023-02034, 0001825034-2023-02041.
 
Event Description
It was reported the patient underwent a procedure approximately 4 years ago.Subsequently, the patient was revised about a month ago due to pain, reactive tissue, fluid in the capsule, and tissue between the baseplate and glenosphere.There is possible metallosis or allergy.Attempts have been made and there is no further information at this time.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
(b)(4).Reported event was unable to be confirmed due to limited information received from the customer.Device history record was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMP AUG MINI BSPLT W TPR SM
Type of Device
EXTREMITIES IMPLANT
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17643707
MDR Text Key322199123
Report Number0001825034-2023-02028
Device Sequence Number1
Product Code KWS
UDI-Device Identifier00880304993235
UDI-Public(01)00880304993235(17)290215(10)64314041
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172502
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number110032410
Device Lot Number64314041
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/04/2023
Initial Date FDA Received08/29/2023
Supplement Dates Manufacturer Received01/29/2024
Supplement Dates FDA Received01/31/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/18/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
SEE H10.
Patient Outcome(s) Required Intervention; Hospitalization;
Patient SexFemale
Patient Weight59 KG
-
-