RESPIRONICS, INC. V60 V60PLUS VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 V60PLUS VENTILATOR |
Device Problem
Blocked Connection (2888)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/18/2023 |
Event Type
malfunction
|
Event Description
|
Philips received a complaint by the customer on the v60, indicating that a "patient circuit is occluded" alarm error occurred.At this time, it is unknown if there was patient involvement at the time the issue was discovered; however, there was no reported harm to the patient or user.An authorized service provider (asp) was dispatched onsite to evaluate and repair the device and confirmed that there was a clamping of an internal tube giving the technical alarm of "patient circuit is occluded"-- the tube was correctly installed, and the device was tested.Following the test procedure and verification for the device, the configuration of the device was restored to the factory settings.The device passed required performance verification tests per philips standard and was returned to service.Additionally, the asp recommended that the customer should use only the accessories and consumables that are approved by philips to guarantee the correct operation of the device.
|
|
Manufacturer Narrative
|
E1: reporting address state: (b)(6).Reporting institution phone #: (b)(6).
|
|
Manufacturer Narrative
|
H10: it is unknown if there was patient involvement at the time the issue was discovered; however, there was no reported harm to the patient or user.An authorized service provider (asp) was dispatched onsite to evaluate and repair the device and confirmed that there was a clamping of an internal tube giving the technical alarm of "patient circuit is occluded"-- the internal pneumatic circuit was checked, and the device was tested.The issue was confirmed.Following the test procedure and verification for the device, the configuration of the device was restored to the factory settings.The device passed required performance verification tests per philips standard and was returned to service.The investigation concludes that no further action is required at this time.If additional information is received, the complaint file will be reopened.Additionally, the asp recommended that the customer should use only the accessories and consumables that are approved by philips to guarantee the correct operation of the device.The device passed required performance verification tests per philips standard and was returned to service.The investigation concludes that no further action is required at this time.If additional information is received, the complaint file will be reopened.
|
|
Search Alerts/Recalls
|
|
|