Catalog Number 306593 |
Device Problem
Contamination (1120)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/11/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the bd pre-filled saline syringe experienced foreign matter in fluid path.The following information was provided by the initial reporter: after the customer got this precharged without removing the outer package, he found black substance inside and near the white cone head cap.
|
|
Manufacturer Narrative
|
H3: a device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
|
Event Description
|
It was reported that the bd pre-filled saline syringe experienced foreign matter in fluid path.The following information was provided by the initial reporter: after the customer got this precharged without removing the outer package, he found black substance inside and near the white cone head cap.
|
|
Manufacturer Narrative
|
The following fields were updated due to additional information: d10: device available for eval yes.D10: returned to manufacturer on: 25aug2023.H6: investigation summary: it was reported there was a black substance inside and near the white cone head cap.To aid in the investigation, one empty sample with an opened packaging flow wrap was received for evaluation by our quality team.The tip cap has embedded discoloration.No other defects or imperfections were observed.This defect could occur if there was contamination from the resin or colorant that became molded into the cap.A device history record review was completed for provided material number 306593, lot 2153613.The review did not reveal any detected quality issues during the production of this lot that could have contributed to the reported defect.There were no related quality notifications.All processes and final inspections complied with specification requirements.The sample will be shown to associates for awareness.To date, there have been no other similar events reported for this lot.Based on the investigation and with the returned sample analysis the symptom reported by the customer is confirmed.H3 other text : see h10.
|
|