• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD MEDICAL (BD WEST) MEDICAL SURGICAL BD PRE-FILLED SALINE SYRINGE; SALINE, VASCULAR ACCESS FLUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL (BD WEST) MEDICAL SURGICAL BD PRE-FILLED SALINE SYRINGE; SALINE, VASCULAR ACCESS FLUSH Back to Search Results
Catalog Number 306593
Device Problem Contamination (1120)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/11/2023
Event Type  malfunction  
Event Description
It was reported that the bd pre-filled saline syringe experienced foreign matter in fluid path.The following information was provided by the initial reporter: after the customer got this precharged without removing the outer package, he found black substance inside and near the white cone head cap.
 
Manufacturer Narrative
H3: a device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that the bd pre-filled saline syringe experienced foreign matter in fluid path.The following information was provided by the initial reporter: after the customer got this precharged without removing the outer package, he found black substance inside and near the white cone head cap.
 
Manufacturer Narrative
The following fields were updated due to additional information: d10: device available for eval yes.D10: returned to manufacturer on: 25aug2023.H6: investigation summary: it was reported there was a black substance inside and near the white cone head cap.To aid in the investigation, one empty sample with an opened packaging flow wrap was received for evaluation by our quality team.The tip cap has embedded discoloration.No other defects or imperfections were observed.This defect could occur if there was contamination from the resin or colorant that became molded into the cap.A device history record review was completed for provided material number 306593, lot 2153613.The review did not reveal any detected quality issues during the production of this lot that could have contributed to the reported defect.There were no related quality notifications.All processes and final inspections complied with specification requirements.The sample will be shown to associates for awareness.To date, there have been no other similar events reported for this lot.Based on the investigation and with the returned sample analysis the symptom reported by the customer is confirmed.H3 other text : see h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PRE-FILLED SALINE SYRINGE
Type of Device
SALINE, VASCULAR ACCESS FLUSH
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer (Section G)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer Contact
jennifer suh
5859 farinon drive
san antonio, TX 78249
8448235433
MDR Report Key17646473
MDR Text Key322255410
Report Number1911916-2023-00621
Device Sequence Number1
Product Code NGT
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number306593
Device Lot Number2153613
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/11/2023
Initial Date FDA Received08/29/2023
Supplement Dates Manufacturer Received08/29/2023
Supplement Dates FDA Received09/13/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/02/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-