Model Number CD2357-40Q
Device Problem Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/10/2023
Event Type  malfunction  
Event Description
It was reported that during procedure while installing a lead a failure to connect to the header was noted on the implantable cardioverter defibrillator (icd).The physician elected to explant and replace the icd.The patient was in stable condition.
 
Manufacturer Narrative
The reported event of lead connection anomaly was confirmed.However, the anomaly was due to the setscrew having been tightened all the way down.Electrical, longevity, and visual analysis was normal with no anomalies found.A device history record (dhr) review was performed and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORTIFY ASSURA DR ICD, US
Type of Device
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17647396
MDR Text Key322313875
Report Number2017865-2023-40635
Device Sequence Number1
Product Code NVZ
UDI-Device Identifier05414734508094
UDI-Public05414734508094
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCD2357-40Q
Device Lot NumberA000141986
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/10/2023
Initial Date FDA Received08/29/2023
Supplement Dates Manufacturer Received01/12/2024
Supplement Dates FDA Received01/17/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/08/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-